CBCT image reconstruction using a mixed Poisson-Gaussian maximum likelihood function

被引:0
|
作者
Zheng R.-Z. [1 ]
Zhao F. [2 ]
Li B. [3 ]
Tian X. [1 ]
机构
[1] School of Electronic Information, Wuhan University, Wuhan
[2] Department of Cardiovascular Disease, Zhongnan Hospital of Wuhan University, Wuhan University, Wuhan
[3] Department of Radiology, Stomatology Hospital of Wuhan University, Wuhan University, Wuhan
关键词
Cone Beam Computed Tomography(CBCT); Image reconstruction; Mixed Poisson-Gaussian maximum likelihood function; Total variation regularization;
D O I
10.3788/OPE.20202802.0457
中图分类号
学科分类号
摘要
Improving the quality of Cone-Beam Computed Tomography (CBCT) reconstructed images under complicated noise conditions was critical for CBCT systems. In this study, an image reconstruction method of CBCT based on a hybrid Poisson-Gaussian maximum likelihood function was proposed. First, a CBCT image reconstruction model suitable for describing a mixed Poisson-Gaussian noise environment was studied. The model contained a fidelity term based on a mixed Poisson-Gaussian maximum likelihood function and a constraint term based on a three-dimensional total variation regularization method. The fidelity term was used to constrain the reconstruction result to match the observed value as closely as possible in the mixed noise model, where as the constraint term was used for noise removal and preserving the edge and detail information as effectively as possible. The proposed model was further solved using the separable approximation and extended Lagrangian methods. Finally, the effectiveness of the algorithm was verified using both simulation and real data. The results indicate that the proposed method exhibited a maximum improvement of 2.1 dB as compared to other methods evaluated usingsimulated data. In visual comparisons, the proposed method demonstrated the best denoising performance. We can thus conclude that the proposed method is effective forCBCT reconstruction under low dose conditions. © 2020, Science Press. All right reserved.
引用
收藏
页码:457 / 464
页数:7
相关论文
共 14 条
  • [1] Lv X.Q., Wu L., Gu Y., Et al., Detection of low dose CT pulmonary nodules based on 3D convolution neural network, Opt. Precision Eng., 26, 5, pp. 1212-1218, (2018)
  • [2] Sun J.Y., Yu C.Y., Dong S.J., Noise reduction of independent component analysis based on NL means noise prediction, Opt. Precision Eng., 26, 2, pp. 512-516, (2018)
  • [3] Yin Y., Yu G., Et al., CBCT Image Denoising based on Multi-scale Wavelet Transform, Proc. of 20103rd International Conference on Biomedical Engineering and Informatics, pp. 16-18, (2010)
  • [4] Hao J., Zhang L., Li L., Kang K., A comparison of projection domain noise reduction methods in low-dose dental CBCT, Proc. of 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference Record, pp. 3624-3627, (2012)
  • [5] Geng L., Ma X., Xiao Z.T., Et al., Calibration and filtering of X-ray line array detector, Infrared and Laser Engineering, 46, 12, pp. 293-300, (2017)
  • [6] Feldkamp L.A., Davis L.C., Kress J.W., Practical cone-beam algorithm, Journal of the Optical Society of America A, 1, 6, pp. 612-619, (1984)
  • [7] Azzari L., Foi A., Variance stabilization in poisson image deblurring, Proc. of 2017 IEEE 14th International Symposium on Biomedical Imaging (ISBI 2017), pp. 728-731, (2017)
  • [8] Zhang Y., Song P., Dai Q., Fourier ptychographic microscopy using a generalized anscombe transform approximation of the mixed poisson-gaussian likelihood, Optics Express, 25, 1, (2017)
  • [9] Zhang B., Fadili M.J., Starck J.L., Marin J.C.O., Multiscale variance-stabilizing transform for mixed-poisson-gaussian processes and its applications in bioimaging, Proc. of 2007 IEEE International Conference on Image Processing, 6, pp. 233-236, (2007)
  • [10] Sidky E.Y., Pan X., Image reconstruction in circular cone-beam computed tomography by constrained, total-variation minimization, Physics in Medicine and Biology, 53, 17, pp. 4777-4807, (2008)