A Stretchable and Wearable Ultrasonic Transducer Array for Bladder Volume Monitoring Application

被引:2
|
作者
Pu C. [1 ]
Fu B. [1 ]
Guo L. [2 ]
Xu H. [3 ]
Peng C. [4 ]
机构
[1] ShanghaiTech University, School of Biomedical Engineering, Shanghai
[2] Shanghai Tenth People's Hospital, Tongji University, Department of Medical Ultrasound, Shanghai
[3] Institute of Ultrasound in Medicine and Engineering, Fudan University, Zhongshan Hospital, Department of Ultrasound, Shanghai
[4] ShanghaiTech University, School of Biomedical Engineering, The State Key Laboratory of Advanced Medical Materials and Devices, Shanghai
关键词
Bladder dysfunction; bladder volume; ultrasonic transducer; wearable sensor; wearable ultrasound;
D O I
10.1109/JSEN.2024.3382244
中图分类号
学科分类号
摘要
Monitoring bladder volume is an essential application for patients with voiding dysfunction. Current wearable ultrasonic bladder monitors are rigid, bulky, and cannot conform to human skin. This study presents a stretchable and wearable ultrasonic transducer array that can both conform to non-planar skin surfaces and continuously monitor bladder volume. The wearable transducer array mainly consists of 4×4 piezoelectric transducer elements, stretchable serpentine-based electrodes for electrical interconnection, electrically conductive adhesive as both matching and backing layers, and silicone elastomer for encapsulation. The transducer array has a center frequency of 3.9 MHz, -6 dB acoustic bandwidth of 57.6%, and up to 40% elastic stretchability. The volume of balloon-bladder phantom was estimated by a least square ellipsoid fitting method. With the true volume of balloon-bladder phantoms ranging from 10 to 405 mL, the proposed transducer array has a mean absolute percentage error (MAPE) of 9.4%, a mean absolute error (MAE) of 11.9 mL, and a coefficient of determination of 0.975, which demonstrates the capability of the proposed stretchable and wearable ultrasonic transducer array for bladder volume monitoring application. © 2001-2012 IEEE.
引用
收藏
页码:15875 / 15883
页数:8
相关论文
共 50 条
  • [1] Design of a Wearable 2-D Ultrasonic Transducer Array for Bladder Measurement
    Joung, Do Yeong
    Jo, Hyeong Geun
    Pyun, Joo Young
    Park, Bum Hun
    Park, Kwan Kyu
    JOURNAL OF THE KOREAN SOCIETY FOR NONDESTRUCTIVE TESTING, 2021, 41 (01) : 39 - 49
  • [2] Design and Optimization of Flexible and Stretchable Ultrasonic Transducer Array for Arterial Blood Pressure Monitoring
    Liu, Chang
    Sun, He
    Ren, Jiahao
    Cheng, Huanyu
    Xie, Mengying
    Liu, Yang
    IEEE SENSORS JOURNAL, 2024, 24 (09) : 15055 - 15064
  • [3] A wearable ultrasonic bladder monitoring device
    Kiczek, Bartlomiej
    Golabek, Michal
    Wojcik, Dariusz
    Kania, Konrad
    Kozlowski, Edward
    Rymarczyk, Tomasz
    Sikora, Jan
    PROCEEDINGS OF THE 2022 THE 28TH ANNUAL INTERNATIONAL CONFERENCE ON MOBILE COMPUTING AND NETWORKING, ACM MOBICOM 2022, 2022, : 886 - 888
  • [4] Flexible and Stretchable Ultrasonic Transducer Array Conformed to Complex Surfaces
    Liu, Wei
    Zhu, Chunling
    Wu, Dawei
    IEEE ELECTRON DEVICE LETTERS, 2021, 42 (02) : 240 - 243
  • [5] Stretchable Composite Acoustic Transducer for Wearable Monitoring of Vital Signs
    Cotur, Yasin
    Kasimatis, Michael
    Kaisti, Matti
    Olenik, Selin
    Georgiou, Charis
    Guder, Firat
    ADVANCED FUNCTIONAL MATERIALS, 2020, 30 (16)
  • [6] Sparse ultrasonic transducer array for structural health monitoring
    Michaels, TE
    Michaels, JE
    REVIEW OF PROGRESS IN QUANTITATIVE NONDESTRUCTIVE EVALUATION, VOLS 23A AND 23B, 2004, 23 : 1468 - 1475
  • [7] Bioinspired Stretchable Transducer for Wearable Continuous Monitoring of Respiratory Patterns in Humans and Animals
    Cotur, Yasin
    Olenik, Selin
    Asfour, Tarek
    Bruyns-Haylett, Michael
    Kasimatis, Michael
    Tanriverdi, Ugur
    Gonzalez-Macia, Laura
    Lee, Hong Seok
    Kozlov, Andrei S.
    Guder, Firat
    ADVANCED MATERIALS, 2022, 34 (33)
  • [8] Highly Sensitive and Stretchable Ultrasonic Transducer Array for Object Internal Characteristics Detection in Robotics
    Zhang, Zhongtan
    Wang, Yancheng
    Mei, Deqing
    Jin, Jie
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2023, 72
  • [9] Continuous Bladder Volume Monitoring System for Wearable Applications
    Shin, Seung-chul
    Moon, Junhyung
    Kye, Saewon
    Lee, Kyoungwoo
    Lee, Yong Seung
    Kang, Hong-Goo
    2017 39TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2017, : 4435 - 4438
  • [10] Flexible ultrasonic transducer array for arterial blood pressure monitoring
    Liu C.
    Sun H.
    Ren J.
    Cheng H.
    Xie M.
    Liu Y.
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2023, 44 (12): : 225 - 234