A METHOD FOR MEASURING THE ABSOLUTE SURFACE SHAPE OF LARGEAPERTURE OPTICAL FLAT

被引:0
|
作者
Zhang, Xudong [1 ]
Xu, Chunfeng [1 ]
Zhang, Leihong [1 ]
Han, Sen [1 ]
Xu, Banglian [1 ]
Zhang, Dawei [1 ]
Jiang, Chenzhe [1 ]
Zhou, Zhixuan [1 ]
Sun, Quan [1 ,2 ]
机构
[1] Univ Shanghai Sci & Technol, Shanghai 200093, Peoples R China
[2] Natl Univ Def Technol, Coll Adv Interdisciplinary Studies, Changsha, Peoples R China
关键词
interferometer; surface shape measurement; absolute measurement; sub-aperture stitching; immune algorithm;
D O I
10.3116/16091833/Ukr.J.Phys.Opt.2024.04022
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The absolute shape measurement of large aperture optical flats usually requires a standard mirror with the same or larger aperture as the element to be measured as the reference surface for calibration. However, the large aperture standard mirror is difficult to process, and the measurement resolution is limited. To solve this problem, a method of measuring the absolute surface shape of a large aperture optical flat is proposed. The aperture optical flat is divided into several sub-apertures using the sub-aperture stitching method. The reference mirror shape is obtained by the odd and even function four-step absolute measurement method and separated to obtain the absolute surface shape of the sub-aperture. The immune algorithm is optimized and applied to the adjacent sub-aperture stitching to obtain the complete absolute surface shape. The 210 mm long flat crystal was measured, and the closed-loop self-test was designed to verify the correctness of the obtained reference mirror shape. The peak and valley value (PV) of reference mirror B is 53.64 nm, and the root mean square (RMS) value is 9.96 nm. The results of the closed-loop self-test are 58.63 nm and 10.86 nm, and the data are basically consistent with the surface shape diagram. The surface shape test results pass the closed-loop self-test. The absolute surface shape results after stitches are compared with those measured by a large aperture interferometer. The deviation of the PV value is less than lambda/135 (lambda=632.8 nm), and the deviation of the RMS value is less than lambda/75. The measurement resolution is 11 times that measured by a large aperture interferometer, which fully proves the accuracy and high resolution of the method.
引用
收藏
页码:4022 / 4037
页数:16
相关论文
共 50 条
  • [1] Absolute measurement of optical flat surface shape based on the conjugate differential method
    Huang, Ya
    Ma, Jun
    Zhu, Rihong
    Yuan, Caojin
    Chen, Lei
    Cai, Huijuan
    Sun, Weiyuan
    OPTICS EXPRESS, 2015, 23 (23): : 29687 - 29697
  • [2] Method for absolute measurement of flat surface
    Yan, Lisong
    Li, Fazhi
    Wang, Xiaokun
    Zhang, Haidong
    Zhang, Lei
    Ma, Donglin
    APPLIED OPTICS, 2020, 59 (33) : 10584 - 10590
  • [3] Four-step shear method for the absolute measurement of a flat surface based on phase measuring deflectometry
    E, Kewei
    Li, Dahai
    Zhang, Chen
    Zhang, Tao
    Li, Mengyang
    Wang, Qin
    Jin, Chengying
    Xiong, Zhao
    APPLIED OPTICS, 2016, 55 (30) : 8419 - 8425
  • [4] The scanning method for measuring surface shape using the regular lattice of the optical vortices
    Fraczek, Ewa
    Mroczka, Janusz
    17TH SLOVAK-CZECH-POLISH OPTICAL CONFERENCE ON WAVE AND QUANTUM ASPECTS OF CONTEMPORARY OPTICS, 2010, 7746
  • [5] TESTING OF AN ABSOLUTE MEASURING FLAT JACK ACCORDING TO THE COMPENSATION METHOD (AWID).
    Kessels, W.
    1985, : 388 - 404
  • [6] A NEW ABSOLUTE METHOD OF MEASURING SURFACE-AREA
    GERASIMOV, GM
    GERSHANIK, AP
    PYATKINA, OP
    MEASUREMENT TECHNIQUES USSR, 1990, 33 (09): : 954 - 958
  • [7] Absolute interferometric testing of a flat surface based on the Laplacian method
    Soltani, Peyman
    Darudi, Ahmad
    OPTICS LETTERS, 2020, 45 (11) : 3026 - 3029
  • [8] ABSOLUTE CALIBRATION OF AN OPTICAL FLAT
    FRITZ, BS
    PROCEEDINGS OF THE SOCIETY OF PHOTO-OPTICAL INSTRUMENTATION ENGINEERS, 1983, 433 : 123 - 130
  • [9] ABSOLUTE CALIBRATION OF AN OPTICAL FLAT
    FRITZ, BS
    OPTICAL ENGINEERING, 1984, 23 (04) : 379 - 383
  • [10] A surface shape test method for a thin flat mirror
    Guo, Yongwei
    OPTIK, 2018, 152 : 116 - 126