Magnetic Property Calibration for Vibrating Sample Magnetometers Based on Magnetic Field Analysis

被引:1
|
作者
Kong, Xiaohan [1 ]
Uehara, Yuji [2 ]
Terauchi, Naoya [3 ]
Sato, Natsuko [3 ]
Matsuda, Yoshibumi [3 ]
Nagano, Masanori [3 ]
Igarashi, Hajime [1 ]
机构
[1] Hokkaido Univ, Grad Sch Informat Sci & Technol, Sapporo, Hokkaido, Japan
[2] Magnet Device Lab Ltd, Kawasaki, Kanagawa, Japan
[3] Taiyo Yuden Co Ltd, Takasaki, Gumma, Japan
关键词
vibrating sample magnetometer; magnetic property; finite element method; DEMAGNETIZING FACTORS;
D O I
10.1109/CEFC61729.2024.10586179
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
this study proposes a method for the calibration of vibrating sample magnetometer (VSM), designed to compensate for the effects of internal demagnetizing field. This method utilizes finite element calculations and evolutionary search to find the optimal demagnetizing factor. This method has been validated to exhibit higher accuracy than existing methods, and the calibrated magnetic property aligns well with the results obtained through the capacitive cancellation method.
引用
收藏
页数:2
相关论文
共 50 条
  • [1] Wide-Range Calibration of Magnetic Moments for Vibrating Sample Magnetometers
    Zhang, Zhigao
    Xu, Zhiyi
    He, Jian
    Lu, Jun
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2022, 71
  • [2] Magnetic Property Calibration of Low-Permeability Materials for Vibrating Sample Magnetometers Using Neural Networks
    Kong, Xiaohan
    Uehara, Yuji
    Terauchi, Naoya
    Sato, Natsuko
    Matsuda, Yoshibumi
    Nagano, Masanori
    Igarashi, Hajime
    IEEE TRANSACTIONS ON MAGNETICS, 2024, 60 (12)
  • [3] MAGNETIC IMAGE EFFECTS IN VIBRATING-SAMPLE MAGNETOMETERS
    STONER, RE
    HERBERT, R
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1969, 14 (03): : 340 - &
  • [4] CALIBRATION OF VIBRATING-SAMPLE MAGNETOMETERS
    CASE, WE
    HARRINGTON, RD
    JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS SECTION C-ENGINEERING AND INSTRUMENTATION, 1966, C 70 (04): : 255 - +
  • [5] THE DIRECT OBSERVATION OF MAGNETIC IMAGES IN ELECTROMAGNET VIBRATING SAMPLE MAGNETOMETERS
    HOON, SR
    WILLCOCK, SNM
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1988, 21 (05): : 480 - 487
  • [6] Calibration of strapdown magnetometers in magnetic field domain
    Gebre-Egziabher, D
    Elkaim, GH
    Powel, JD
    Parkinson, BW
    JOURNAL OF AEROSPACE ENGINEERING, 2006, 19 (02) : 87 - 102
  • [7] Calibration of strapdown magnetometers in magnetic field domain
    Gebre-Egziabher, Demoz
    Elkaim, Gabriel H.
    Powell, J. David
    Parkinson, Bradford W.
    Journal of Aerospace Engineering, 2006, 19 (02): : 87 - 102
  • [8] Calibration of vibrating sample magnetometers independent of reference samples
    Ausserlechner, U
    Steiner, W
    Kasperkovitz, P
    MEASUREMENT SCIENCE AND TECHNOLOGY, 1996, 7 (11) : 1574 - 1578
  • [9] ON THE CALIBRATION OF VIBRATING SAMPLE MAGNETOMETERS WITH THE HELP OF NICKEL REFERENCE SAMPLES
    ECKERT, D
    SIEVERT, J
    IEEE TRANSACTIONS ON MAGNETICS, 1993, 29 (06) : 3001 - 3003
  • [10] PICK-UP SYSTEMS FOR VIBRATING SAMPLE MAGNETOMETERS - A THEORETICAL DISCUSSION BASED ON MAGNETIC MULTIPOLE EXPANSIONS
    AUSSERLECHNER, U
    KASPERKOVITZ, P
    STEINER, W
    MEASUREMENT SCIENCE AND TECHNOLOGY, 1994, 5 (03) : 213 - 225