Deep learning-aided respiratory motion compensation in PET/CT: addressing motion induced resolution loss, attenuation correction artifacts and PET-CT misalignment

被引:3
|
作者
Lu, Yihuan [1 ]
Kang, Fei [2 ]
Zhang, Duo [1 ]
Li, Yue [1 ]
Liu, Hao [1 ]
Sun, Chen [1 ]
Zeng, Hao [1 ]
Shi, Lei [1 ]
Zhao, Yumo [1 ]
Wang, Jing [2 ]
机构
[1] United Imaging Healthcare, 2258 Chengbei Rd, Shanghai 201807, Peoples R China
[2] Fourth Mil Med Univ, Xijing Hosp, Dept Nucl Med, 127 West Changle Rd, Xian 710032, Shaanxi, Peoples R China
基金
中国国家自然科学基金;
关键词
PET; Data-driven; Respiratory motion correction; uExcel OncoFocus; Deep learning; uMI Panorama; RECONSTRUCTION; PROTOCOL;
D O I
10.1007/s00259-024-06872-x
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeRespiratory motion (RM) significantly impacts image quality in thoracoabdominal PET/CT imaging. This study introduces a unified data-driven respiratory motion correction (uRMC) method, utilizing deep learning neural networks, to solve all the major issues caused by RM, i.e., PET resolution loss, attenuation correction artifacts, and PET-CT misalignment.MethodsIn a retrospective study, 737 patients underwent [18F]FDG PET/CT scans using the uMI Panorama PET/CT scanner. Ninety-nine patients, who also had respiration monitoring device (VSM), formed the validation set. The remaining data of the 638 patients were used to train neural networks used in the uRMC. The uRMC primarily consists of three key components: (1) data-driven respiratory signal extraction, (2) attenuation map generation, and (3) PET-CT alignment. SUV metrics were calculated within 906 lesions for three approaches, i.e., data-driven uRMC (proposed), VSM-based uRMC, and OSEM without motion correction (NMC). RM magnitude of major organs were estimated.ResultsuRMC enhanced diagnostic capabilities by revealing previously undetected lesions, sharpening lesion contours, increasing SUV values, and improving PET-CT alignment. Compared to NMC, uRMC showed increases of 10% and 17% in SUVmax and SUVmean across 906 lesions. Sub-group analysis showed significant SUV increases in small and medium-sized lesions with uRMC. Minor differences were found between VSM-based and data-driven uRMC methods, with the SUVmax was found statistically marginal significant or insignificant between the two methods. The study observed varied motion amplitudes in major organs, typically ranging from 10 to 20 mm.ConclusionA data-driven solution for respiratory motion in PET/CT has been developed, validated and evaluated. To the best of our knowledge, this is the first unified solution that compensates for the motion blur within PET, the attenuation mismatch artifacts caused by PET-CT misalignment, and the misalignment between PET and CT.
引用
收藏
页码:62 / 73
页数:12
相关论文
共 50 条
  • [1] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT
    Osman, MM
    Cohade, C
    Nakamoto, Y
    Wahl, RL
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (04) : 603 - 606
  • [2] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT
    Medhat M. Osman
    Christian Cohade
    Yuji Nakamoto
    Richard L. Wahl
    European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : 603 - 606
  • [3] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT.
    Osman, MM
    Cohade, C
    Wahl, RL
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 305P - 305P
  • [4] Reduction of attenuation correction artifacts in PET-CT
    Nuyts, Johan
    Stroobants, Sigrid
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1895 - 1899
  • [5] Respiratory motion artefacts in pet imaging with CT attenuation correction
    Vale, J.
    Costa, D.
    Fonseca, G.
    Patrina, J.
    Faria, D.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S272 - S272
  • [6] Respiratory motion artefacts in PET imaging with CT attenuation correction
    Vale, J.
    Costa, D. C.
    Fonseca, G.
    Patrina, J.
    Faria, D.
    Ponte, F.
    Teixeira, J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 : S355 - S355
  • [7] Respiratory Motion Correction Strategies in Thoracic PET-CT Imaging
    Nehmeh, Sadek A.
    PET CLINICS, 2013, 8 (01) : 29 - +
  • [8] Challenges of Attenuation Correction in Conventional Respiratory Motion Corrected PET/CT
    Mawlawi, O.
    Meier, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E552 - E552
  • [9] Respiratory motion correction in PET with gated PET/CT data
    Iglesias, J. E.
    Kimiaei, S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 : S315 - S315
  • [10] A Generic Respiratory Motion Model for Motion Correction in PET/CT
    Fayad, H.
    Pan, T.
    Roux, C.
    Visvikis, D.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2455 - 2458