A hybrid PCA acceleration method for rapid real-time 2D MRI

被引:0
|
作者
Wright, Mark [1 ]
Han, Gawon [1 ]
Yun, Jihyun [1 ,2 ]
Yip, Eugene [1 ,2 ]
Gabos, Zsolt [1 ]
Usmani, Nawaid [1 ]
Fallone, B. Gino [1 ,2 ]
Wachowicz, Keith [1 ,2 ]
机构
[1] Univ Alberta, Cross Canc Inst, Dept Oncol, Edmonton, AB, Canada
[2] Cross Canc Inst, Dept Med Phys, Edmonton, AB, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2024年 / 69卷 / 20期
关键词
real-time; MRI; Linac-MR; reconstruction; PCA; LINAC; MOTION;
D O I
10.1088/1361-6560/ad7dcb
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. To develop a 2D MR acceleration method utilizing principal component analysis (PCA) in a hybrid fashion for rapid real-time applications. Approach. Retrospective testing was performed on 10 lung, 10 liver and 10 prostate 3T MRI data sets for image quality and target contourability. Sampling of k-space is performed by acquiring central (low-frequency) data in every frame while the high-frequency data is incoherently undersampled such that all of k-space is acquired in a pre-determined number of frames. Firstly, principal components (PCs) representative of intra-frame correlations between central and outer k-space data are used to estimate unsampled data in the frame of interest. Then to add further stability, PCs representative of time-domain fluctuations within a reconstruction window of the most recent frames are fit to outer k-space data (including above estimations) to obtain final estimates in the frame of interest. Accelerated reconstructions between 3x and 8x were tested for image quality and contourability along with the optimal number of PCs for fitting. Main results. It was found that at higher acceleration rates, image quality did not deteriorate significantly. Similarly, it was found that the images were of sufficient quality to contour a target using auto-contouring software at all tested acceleration rates and sites. SSIM values were found to be >= 0.91 at all accelerations tested. Similarly dice coefficients at the different sites were found to be >= 0.89 even at 8x accelerations which is on par with or better than intra-observer variation. Significance. This method appears to produce improved image quality and contourability compared to previous PCA methods while also allowing a greater number of PCs to be used in reconstruction. The method can be run using a simple single-channel coil and does not require significant computing power to meet real-time interventional standards (reconstruction times similar to 60 ms/frame on Intel i5 CPU).
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Time domain principal component analysis for rapid, real-time 2D MRI reconstruction from undersampled data
    Wright, Mark
    Dietz, Bryson
    Yip, Eugene
    Yun, Jihyun
    Gabos, Zsolt
    Fallone, B. Gino
    Wachowicz, Keith
    MEDICAL PHYSICS, 2021, 48 (11) : 6724 - 6739
  • [2] Extraction of 3D trajectories of mandibular condyles from 2D real-time MRI
    Isaieva, Karyna
    Leclere, Justine
    Paillart, Guillaume
    Drouot, Guillaume
    Felblinger, Jacques
    Dubernard, Xavier
    Vuissoz, Pierre-Andre
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2025, 38 (01): : 131 - 140
  • [3] A Robust Real-Time Acceleration and Reconstruction Scheme for Rapid MRI Using Principal Component Analysis
    Wright, M.
    Dietz, B.
    Yip, E.
    Yun, J.
    Fallone, B. G.
    Wachowicz, K.
    MEDICAL PHYSICS, 2020, 47 (06) : E510 - E511
  • [4] ImageDosis: 2D real-time dosimetry system
    Nascimento, L. De Freitas
    Verellen, D.
    Goossens, J.
    Vanhavere, F.
    Akselrod, M.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S354 - S354
  • [5] Real-Time Motion Tracking and Prediction Using 2D Cine MRI for Abdominal Mrgrt
    Tseng, W. C.
    Perez-Sanchez, J. A.
    Sun, S. H.
    Liu, C.
    Yan, G.
    MEDICAL PHYSICS, 2024, 51 (09) : 6586 - 6586
  • [6] Three-dimensional liver motion tracking using 2D real-time MRI
    Brix, L.
    Ringgaard, S.
    Sorensen, T. Sangild
    Poulsen, P. Rugaard
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S86 - S86
  • [7] Multimodal dataset of real-time 2D and static 3D MRI of healthy French speakers
    Karyna Isaieva
    Yves Laprie
    Justine Leclère
    Ioannis K. Douros
    Jacques Felblinger
    Pierre-André Vuissoz
    Scientific Data, 8
  • [8] Multimodal dataset of real-time 2D and static 3D MRI of healthy French speakers
    Isaieva, Karyna
    Laprie, Yves
    Leclere, Justine
    Douros, Ioannis K.
    Felblinger, Jacques
    Vuissoz, Pierre-Andre
    SCIENTIFIC DATA, 2021, 8 (01)
  • [9] Real-time 2-D fluid simulation with a GPU acceleration
    School of Mechactronics Engineering, Harbin Institute of Technology, Harbin 150001, China
    Harbin Gongcheng Daxue Xuebao, 2008, 3 (278-284): : 278 - 284
  • [10] Real-time 2D to 3D video conversion
    Ianir Ideses
    Leonid P. Yaroslavsky
    Barak Fishbain
    Journal of Real-Time Image Processing, 2007, 2 : 3 - 9