Isotopic analysis of uranium at the picogram level with single-shot laser induced breakdown spectroscopy
被引:4
|作者:
Russo, Richard E.
[1
,2
]
Gonzalez, Jhanis J.
[1
]
Oropeza, Dayana
[2
]
Liu, Chunyi
[1
]
Chirinos, Jose
[2
]
Chan, George C. -Y.
[2
]
机构:
[1] Appl Spectra, 950 Riverside Pkwy Suite 90, West Sacramento, CA 95605 USA
[2] Lawrence Berkeley Natl Lab, Berkeley, CA 94720 USA
关键词:
Isotopic analysis;
Isotopic shift;
Laser induced breakdown spectroscopy;
Nuclear analysis;
Particle analysis;
NEUTRAL-URANIUM;
PLASMA;
LIBS;
PARTICLES;
ABSORPTION;
PARAMETERS;
PRECISION;
D O I:
10.1016/j.sab.2024.106928
中图分类号:
O433 [光谱学];
学科分类号:
0703 ;
070302 ;
摘要:
Uranium was measured by laser induced breakdown spectroscopy (LIBS) with a limit of detection of 1.3 picograms with spectral-peak height as the analytical parameter, which shows an improvement of several orders of magnitude compared to other reports in the literature. The improvement is a result of optimizing the choice of the U analytical line, the system hardware, and the light collection and detection efficiency. The research established properties of U line emission (intensity, signal-to-background ratio, isotopic splitting, and spectral interferences) to select an optimum wavelength for analysis. The samples were mock particles from dried, monodispersed U solution droplets (similar to 65 pL in volume) deposited on Si wafers. LIBS emission spectra from several U-235-enriched samples were measured to demonstrate the performance for isotopic analysis. For isotopic measurements, two masses for total U, 100 pg and 325 pg, were studied. The U-235: U-238 isotopic ratio was obtained by peak fitting the LIBS spectra without requiring calibration with the use of enriched standards. Depending on the total U mass and the U-235 enrichment, typical relative standard deviations of the isotopic analysis ranged from several percent down to about 1%. Further, the developed method and the reported analytical performance were all based on single-shot LIBS. Overall, the study highlights the potential of LIBS as a robust and sensitive technique for isotopic analysis of U-containing particles.
引用
收藏
页数:9
相关论文