Dosimetric benefits and preclinical performance of steerable needles in HDR prostate brachytherapy

被引:1
|
作者
de Vries, M. [1 ]
Christianen, M. E. M. C. [2 ]
Luthart, L. [2 ]
de Vries, K. C. [2 ]
Kolkman-Deurloo, I. K. K. [2 ]
van den Dobbelsteen, J. J. [1 ]
机构
[1] Delft Univ Technol, Fac Mech Maritime & Mat Engn, Dept Biomech Engn, Mekelweg 2, NL-2628 CD Delft, Netherlands
[2] Univ Med Ctr Rotterdam, Erasmus MC Canc Inst, Dept Radiotherapy, Dr Molewaterpl 40, NL-3015 GD Rotterdam, Netherlands
关键词
HDR; Brachytherapy; Prostate cancer; Pubic arch interference; Steerable needle; PUBIC ARCH INTERFERENCE; DOSE-RATE BRACHYTHERAPY; AMERICAN BRACHYTHERAPY; CANCER; FEASIBILITY; MONOTHERAPY; VOLUME; MEN;
D O I
10.1016/j.medengphy.2024.104177
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Prostate cancer patients with an enlarged prostate and/or excessive pubic arch interference (PAI) are generally considered non-eligible for high-dose-rate (HDR) brachytherapy (BT). Steerable needles have been developed to make these patients eligible again. This study aims to validate the dosimetric impact and performance of steerable needles within the conventional clinical setting. HDR BT treatment plans were generated, needle implantations were performed in a prostate phantom, with prostate volume > 55 cm(3) and excessive PAI of 10 mm, and pre- and post-implant dosimetry were compared considering the dosimetric constraints: prostate V-100 > 95 % (13.50 Gy), urethra D-0.1cm(3) < 115 % (15.53 Gy) and rectum D-1cm(3) < 75 % (10.13 Gy). The inclusion of steerable needles resulted in a notable enhancement of the dose distribution and prostate V-100 compared to treatment plans exclusively employing rigid needles to address PAI. Furthermore, the steerable needle plan demonstrated better agreement between pre- and post-implant dosimetry (prostate V-100: 96.24 % vs. 93.74 %) compared to the rigid needle plans (79.13 % vs. 72.86 % and 87.70 % vs. 81.76 %), with no major changes in the clinical workflow and no changes in the clinical set-up. The steerable needle approach allows for more flexibility in needle positioning, ensuring a highly conformal dose distribution, and hence, HDR BT is a feasible treatment option again for prostate cancer patients with an enlarged prostate and/or excessive PAI.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] MR-guided HDR prostate brachytherapy with teleoperated steerable needles
    M. de Vries
    M. Wijntjes
    J. Sikorski
    P. Moreira
    N. J. van de Berg
    J. J. van den Dobbelsteen
    S. Misra
    Journal of Robotic Surgery, 2023, 17 : 2461 - 2469
  • [2] MR-guided HDR prostate brachytherapy with teleoperated steerable needles
    de Vries, M.
    Wijntjes, M.
    Sikorski, J.
    Moreira, P.
    van de Berg, N. J.
    van den Dobbelsteen, J. J.
    Misra, S.
    JOURNAL OF ROBOTIC SURGERY, 2023, 17 (05) : 2461 - 2469
  • [3] The dosimetric impact of interstitial needles in HDR brachytherapy for cervical cancer
    Otter, S.
    Coates, A.
    Franklin, A.
    Brennan, C.
    Cunningham, M.
    Stewart, A.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127
  • [4] Dosimetric impact of uncertainty in the reconstruction of needles in HDR prostate
    Colmenares, R.
    Polo, A.
    Hervas, A.
    Moris, R.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S564 - S565
  • [5] HDR Prostate Brachytherapy - Is 20 needles too many?
    Wilson, J.
    Lydon, A.
    Rowberry, L.
    Trewin, C.
    RADIOTHERAPY AND ONCOLOGY, 2021, 158 : S199 - S199
  • [6] A new set of needles and matrixes for HDR brachytherapy of prostate cancer
    Lundell, M
    Nilsson, J
    Nilsson, S
    Hellberg, B
    Inácio, J
    Persson, S
    Ryberg, M
    Kälkner, KM
    RADIOTHERAPY AND ONCOLOGY, 2005, 75 : S37 - S38
  • [7] An automated optimization tool for HDR prostate brachytherapy with divergent needles
    de Battisti, M. Borot
    Maenhout, M.
    de Senneville, B. Denis
    Hautvast, G.
    Binnekamp, D.
    Lagendijk, J. J. W.
    Van Vulpen, M.
    Moerland, M. A.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S88 - S89
  • [8] Nomogram as an Efficient Dosimetric Verification Tool in HDR Prostate Brachytherapy
    Liang, J.
    Todor, D.
    MEDICAL PHYSICS, 2016, 43 (06) : 3626 - 3626
  • [9] Steerable needles enable HDR prostate BT for patients with pubic arch interference
    de Vries, M.
    Christianen, M.
    Luthart, L.
    de Vries, K.
    Kolkman-Deurloo, I.
    van den Dobbelsteen, J.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S315 - S317
  • [10] Overcoming pubic arch interference in prostate brachytherapy using steerable needles
    de Vries, Martijn
    Wilby, Sarah L.
    Palmer, Antony L.
    Polak, Wojciech
    O'Hea, Inna
    Hodgson, Dominic
    van den Dobbelsteen, John J.
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2022, 14 (05) : 495 - 500