Retrospective Tube Current Modulation Optimization of Individualized Organ-Level CT Dose and Image Quality

被引:1
|
作者
Wang, Sen [1 ]
Medrano, Maria [1 ]
Imran, Abdullah-Al-Zubaer [1 ]
Stevens, Grant [3 ]
Tse, Justin Ruey [1 ]
Wang, Adam S. [1 ,2 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
[3] GE HealthCare, Waukesha, WI 53188 USA
关键词
Tube current modulation; computed tomography; image quality; organ dose; REDUCTION;
D O I
10.1117/12.3006870
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Automatic exposure control based on tube current modulation (TCM) can effectively reduce dose while maintaining image quality. Conventional TCM uses total exposure from the tube and noise in the center of CT slices as surrogates of dose and image quality, respectively. In this abstract, we present an automated method to optimize TCM at the organ level, offering increased flexibility and aligning with the concept of organ-specific radiation risk assessment. We applied our method to a retrospective CT dataset and incorporated automatic organ segmentation, Monte Carlo simulation for dose calculation, and an empirical model for noise estimation. This method was fully automated and readily scalable to massive clinical data, allowing the generation of ground-truth data for any data-driven approach to prospective planning, including methods utilizing scout images.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Prospective Optimization of CT under Tube Current Modulation: I. Organ Dose
    Tian, Xiaoyu
    Li, Xiang
    Segars, W. Paul
    Frush, Donald P.
    Samei, Ehsan
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [2] COMPARISON OF ORGAN-BASED TUBE CURRENT MODULATION AND BISMUTH SHIELDING IN CHEST CT: EFFECT ON THE IMAGE QUALITY AND THE PATIENT DOSE
    Kotiaho, Antti
    Manninen, Anna-Leena
    Nikkinen, Juha
    Nieminen, Miika Tapio
    RADIATION PROTECTION DOSIMETRY, 2019, 185 (01) : 50 - 56
  • [3] Prospective estimation of organ dose in CT under tube current modulation
    Tian, Xiaoyu
    Li, Xiang
    Segars, W. Paul
    Frush, Donald P.
    Samei, Ehsan
    MEDICAL PHYSICS, 2015, 42 (04) : 1575 - 1585
  • [4] Implementation of Tube Current Modulation into An Organ Dose Calculator for CT Patients
    Lee, C.
    Yeom, Y.
    Atkinson, J.
    Folio, L.
    MEDICAL PHYSICS, 2020, 47 (06) : E264 - E264
  • [5] Prospective Optimization of CT under Tube Current Modulation: II. Image Quality
    Tian, Xiaoyu
    Wilson, Josh
    Frush, Donald
    Samei, Ehsan
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [6] Impact of head CT organ dose modulation on lens dose and image quality
    Dalah, Entesar Z.
    Grazyna, Wasilewska
    Al Bastaki, Usama M.
    Abedzadeh, Ayoub A.
    AL Musfari, Yousuf M.
    Khan, Sabaa A.
    Elawad, Noon M.
    Bradley, David A.
    RADIATION PHYSICS AND CHEMISTRY, 2024, 223
  • [7] Head CT dose reduction with organ-based tube current modulation
    Ota, Joji
    Yokota, Hajime
    Kobayashi, Takenori
    Ogata, Yuki
    Kubo, Takumi
    Chida, Koichi
    Masuda, Yoshitada
    Uno, Takashi
    MEDICAL PHYSICS, 2022, 49 (03) : 1964 - 1971
  • [8] Effective dose and image quality evaluations of an automatic CT tube current modulation system with an anthropomorphic phantom
    Rampado, O.
    Marchisio, F.
    Izzo, A.
    Garelli, E.
    Bianchi, C. C.
    Gandini, G.
    Ropolo, R.
    EUROPEAN JOURNAL OF RADIOLOGY, 2009, 72 (01) : 181 - 187
  • [9] Comparison of Image Quality and Radiation Dose between Fixed Tube Current and Combined Automatic Tube Current Modulation in Craniocervical CT Angiography
    Lee, E. J.
    Lee, S. K.
    Agid, R.
    Howard, P.
    Bae, J. M.
    terBrugge, K.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2009, 30 (09) : 1754 - 1759
  • [10] Image quality and age-specific dose estimation in head and chest CT examinations with organ-based tube-current modulation
    Yamauchi-Kawaura, C.
    Yamauchi, M.
    Imai, K.
    Ikeda, M.
    Aoyama, T.
    RADIATION PROTECTION DOSIMETRY, 2013, 157 (02) : 193 - 205