Optimizing Microfluidic Impedance Cytometry by Bypass Electrode Layout Design

被引:1
|
作者
Wu, Guangzu [1 ,2 ]
Zhang, Zhiwei [1 ,2 ]
Du, Manman [3 ]
Wu, Dan [4 ]
Zhou, Junting [4 ]
Hao, Tianteng [4 ]
Xie, Xinwu [1 ,2 ]
机构
[1] Acad Mil Sci, Peoples Liberat Army, Syst Engn Inst, Tianjin 300161, Peoples R China
[2] Natl BioProtect Engn Ctr, Tianjin 300161, Peoples R China
[3] Tianjin Univ, Sch Environm Sci & Engn, Tianjin 300072, Peoples R China
[4] Tianjin Univ Sci & Technol, Sch Elect Informat & Automat, Tianjin 300222, Peoples R China
来源
BIOSENSORS-BASEL | 2024年 / 14卷 / 04期
关键词
microfluidic impedance cytometry; no bypass electrode; floating electrode; grounding electrode; sensing sensitivity;
D O I
10.3390/bios14040204
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Microfluidic impedance cytometry (MIC) has emerged as a popular technique for single-cell analysis. Traditional MIC electrode designs consist of a pair of (or three) working electrodes, and their detection performance needs further improvements for microorganisms. In this study, we designed an 8-electrode MIC device in which the center pair was defined as the working electrode, and the connection status of bypass electrodes could be changed. This allowed us to compare the performance of layouts with no bypasses and those with floating or grounding electrodes by simulation and experiment. The results of detecting Phi 5 mu m beads revealed that both the grounding and the floating electrode outperformed the no bypass electrode, and the grounding electrode demonstrated the best signal-to-noise ratio (SNR), coefficient of variation (CV), and detection sensitivity. Furthermore, the effects of different bypass grounding areas (numbers of grounding electrodes) were investigated. Finally, particles passing at high horizontal positions can be detected, and Phi 1 mu m beads can be measured in a wide channel (150 mu m) using a fully grounding electrode, with the sensitivity of bead volume detection reaching 0.00097%. This provides a general MIC electrode optimization technology for detecting smaller particles, even macromolecular proteins, viruses, and exosomes in the future.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Coplanar electrode microfluidic chip enabling accurate sheathless impedance cytometry
    De Ninno, Adele
    Errico, Vito
    Bertani, Francesca Romana
    Businaro, Luca
    Bisegna, Paolo
    Caselli, Federica
    LAB ON A CHIP, 2017, 17 (06) : 1158 - 1166
  • [2] Modified trident-shaped electrode design for particle lateral position detection in microfluidic impedance flow cytometry
    Peng, Yunhao
    Gale, Bruce K.
    Porter, Marc D.
    Zang, Ling
    Sant, Himanshu J.
    SENSORS AND ACTUATORS A-PHYSICAL, 2025, 381
  • [3] Microfluidic impedance cytometry for single-cell sensing: Review on electrode configurations
    Zhu, Shu
    Zhang, Xiaozhe
    Zhou, Zheng
    Han, Yu
    Xiang, Nan
    Ni, Zhonghua
    TALANTA, 2021, 233
  • [4] Development of a Novel Design of Microfluidic Impedance Cytometry for Improved Sensitivity and Cell Identification
    Warren, Michael. A.
    Shakouri, Amir
    Pacheco-Pena, Victor
    Hallam, Toby
    ACS OMEGA, 2023, 8 (21): : 18882 - 18890
  • [5] Microfluidic impedance cytometry of tumour cells in blood
    Spencer, Daniel
    Hollis, Veronica
    Morgan, Hywel
    BIOMICROFLUIDICS, 2014, 8 (06):
  • [6] Microfluidic impedance cytometry for blood cell analysis
    Morgan, Hywel
    Spencer, Daniel
    RSC Nanoscience and Nanotechnology, 2015, 2015-January (36): : 213 - 241
  • [7] Microfluidic Impedance-Based Flow Cytometry
    Cheung, Karen C.
    Di Berardino, Marco
    Schade-Kampmann, Grit
    Hebeisen, Monika
    Pierzchalski, Arkadiusz
    Bocsi, Jozsef
    Mittag, Anja
    Tarnok, Attila
    CYTOMETRY PART A, 2010, 77A (07) : 648 - 666
  • [8] Design of piezoelectric modal filters by simultaneously optimizing the structure layout and the electrode profile
    Ruiz, D.
    Bellido, J. C.
    Donoso, A.
    STRUCTURAL AND MULTIDISCIPLINARY OPTIMIZATION, 2016, 53 (04) : 715 - 730
  • [9] Design of piezoelectric modal filters by simultaneously optimizing the structure layout and the electrode profile
    D. Ruiz
    J. C. Bellido
    A. Donoso
    Structural and Multidisciplinary Optimization, 2016, 53 : 715 - 730
  • [10] An optimizing design method for electrode structure parameter of electrical impedance tomography
    Wang Yan
    Sha Hong
    Ren Chaoshi
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 1525 - 1528