Cardiac Migration of Chemoport Catheter-A Rare Complication: Review of Literature and Our Experience

被引:0
|
作者
Dave, Samarth [1 ]
Khunteta, Nitin [1 ]
Mohan, Anand [1 ]
Agarwal, Deepesh [2 ]
Das, Mrinal [1 ]
Viswanath, Mohinder [1 ]
Yadav, Dinesh [1 ]
Sharma, Sanjay [1 ]
Sharma, Rajgovind [1 ]
机构
[1] MGUMST, Mahatma Gandhi Med Coll & Hosp, Dept Surg Oncol, SRCC, Jaipur, Rajasthan, India
[2] MGUMST, Mahatma Gandhi Med Coll & Hosp, Dept Cardiol, SRCC, Jaipur, Rajasthan, India
关键词
Chemoport; Cardiac migration; Pinch-off syndrome; PINCH-OFF SYNDROME; PERCUTANEOUS RETRIEVAL;
D O I
10.1007/s13193-024-01949-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Chemoport devices are commonly used for the administration of chemotherapeutic medication in cancer patients and can also be used for total parenteral nutrition and long term intravenous therapy. They are usually inserted through the internal jugular vein or the subclavian vein. The common complications of the procedure include pneumothorax, bleeding, arrhythmia and venous thrombosis. One of the rare complications of chemoport is catheter fracture/dislodgement and subsequent migration, with an incidence of 0.1-2.1%. Other rare complications are vascular erosion and embolization, vocal cord palsy and mediastinal hematoma. The aims and objectives are as follows: (1) to report a rare case of chemoport catheter migration between the right ventricle and pulmonary artery and (2) to review the literature on the rare complication of cardiac migration of a chemoport. A 56-year-old lady, known case of carcinoma right breast, post modified radical mastectomy was advised adjuvant chemotherapy. A chemoport catheter was placed in the right internal jugular vein and was positioned over the left upper chest. The 1st cycle of chemotherapy was given through chemoport and was uneventful. On the second chemotherapy schedule, catheter dysfunction was found. For the same, she was evaluated with chest radiography, which showed the migrated catheter in the heart. The migrated catheter was retrieved by snare technique using percutaneous transvenous route. The procedure was uneventful. The literature on the topic was reviewed. Chemoport catheter fracture or dislodgement and its subsequent cardiac migration are a rare but serious complication. High index of suspicion in case of catheter dysfunction, early detection by chest radiography, and timely multi-disciplinary intervention is crucial.
引用
收藏
页码:590 / 593
页数:4
相关论文
共 50 条
  • [1] Intracardiac migration of distal catheter-a rare complication of VP shunt insertion: case report and literature review
    Hobbs, Ella
    Thompson, Dominic N. P.
    Muthialu, Nagarajan
    Silva, Adikarige Haritha Dulanka
    CHILDS NERVOUS SYSTEM, 2024, 40 (02) : 587 - 591
  • [2] Migration of chemoport catheter to the right ventricle: A catastrophic rare complication
    Goyal, Neha
    Kamal, Manoj
    Paliwal, Bharat
    Kumar, Rakesh
    SAUDI JOURNAL OF ANAESTHESIA, 2022, 16 (01) : 124 - +
  • [3] Intraluminal Stone in a PD Catheter-A Rare Complication
    Skaro, D. Boric
    Jelicic, I.
    Ljutic, D.
    PERITONEAL DIALYSIS INTERNATIONAL, 2011, 31 (03): : 371 - 372
  • [4] Retained Catheter Fragment from a Fractured Tunneled Catheter-A Rare and Potentially Lethal Complication
    Reddy, Anand
    Stangl, Anondo
    Radbill, Brian
    SEMINARS IN DIALYSIS, 2010, 23 (05) : 536 - 539
  • [5] Intramuscular Migration of Venous Catheter as a Rare Complication of Ventriculoatrial Shunt: Case Report and Literature Review
    Hung, Cheng-Che
    Chuang, Hao-Yu
    Lin, Hung-Lin
    Chu, Yen-Tse
    Cheng, Cheng Hsin
    JOURNAL OF NEUROLOGICAL SURGERY PART A-CENTRAL EUROPEAN NEUROSURGERY, 2017, 78 (04) : 412 - 416
  • [6] Intracardiac migration of distal catheter—a rare complication of VP shunt insertion: case report and literature review
    Ella Hobbs
    Dominic N. P. Thompson
    Nagarajan Muthialu
    Adikarige Haritha Dulanka Silva
    Child's Nervous System, 2024, 40 : 587 - 591
  • [7] The stuck haemodialysis catheter-a case report of a rare but dreaded complication following kidney transplantation
    Burnett, Cameron
    Chandler, S.
    Jegatheesan, D.
    Pearch, B.
    Viecelli, A.
    Mudge, D. W.
    BMC NEPHROLOGY, 2024, 25 (01)
  • [8] Catheter fracture and cardiac migration: A rare complication of totally implantable venous devices
    Di Carlo, I
    Fisichella, P
    Russello, D
    Puleo, S
    Latteri, F
    JOURNAL OF SURGICAL ONCOLOGY, 2000, 73 (03) : 172 - 173
  • [9] Rare complication of ventriculoperitoneal shunt. Earlyonsetofdistal catheter migration into scrotum in an adult male: Casereportand literature review
    Lee, Bryan S.
    Vadera, Sumeet
    Gonzalez-Martinez, Jorge A.
    INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS, 2015, 6 : 198 - 202
  • [10] Chemoport Fracture due to Catheter Pinch Off Syndrome: A Rare Complication of Subclavian Vein Approach Revisited
    Harish Neelamraju Lakshmi
    Mohit Sharma
    Abhishek Jain
    Ketul Puj
    Dipin Jayaprakash
    Shashank Pandya