Calibration method of the right-angle error of a hollow corner-cube retroreflector based on an independent autocollimator

被引:1
|
作者
Li, Renpu [1 ]
Huo, Yujia [1 ]
Yan, Juan [1 ]
Wen, Dandan [1 ]
Konyakhin, Igor [2 ]
Dang, Dinhduan [2 ]
Zhou, Xingye [3 ]
Huang, Guifu [1 ]
Ma, Yong [1 ]
机构
[1] Chongqing Univ Posts & Telecommun, Sch Optoelect Engn, Chongqing 400065, Peoples R China
[2] ITMO Univ, Higher Sch Engn & Technol, 49 Kronverkskiy Prospect, St Petersburg 197101, Russia
[3] Hebei Semicond Res Inst, Natl Key Lab Applicat Specif Integrated Circuit, Shijiazhuang 050051, Peoples R China
基金
中国国家自然科学基金;
关键词
Compendex;
D O I
10.1364/AO.510872
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Hollow corner -cube retroreflectors (HCCRs) are an essential reflection component of next -generation lunar laser -ranging technology. The verticality among the three reflectors, known as the right-angle error, is a critical parameter that affects the emission performance, and thus, should be correctly measured and calibrated. However, conventional methods measure the three right-angle errors separately, which can induce error superposition during the measurement process. A one-time measurement method was developed for the three right-angle errors of the HCCR using a single autocollimator (AC). The method establishes a mathematical relationship between the rightangle error of the HCCR and the angle offset of the reflected beam, and it considers the observation coordinates of the AC simultaneously to perform the coordinate transformation of the relationship parameters. The corresponding measurement equation was derived to extract the three -plane right-angle error of the HCCR using the measured readings of a single AC. In addition, a HCCR was designed to freely adjust the angle of the two reflective surfaces and used to simulate the different states of the three right-angle errors in practice. The measurement results show that the root -mean -square error of the proposed method in all right-angle error states is smaller than 1600. (c) 2024 Optica Publishing Group
引用
收藏
页码:668 / 675
页数:8
相关论文
共 25 条
  • [1] Roll angle of autocollimator measurement method based on hollow cube corner reflector
    Liu Yu
    Zhen Yen
    Xie Linyi
    Wang Wei
    Yang Jiawen
    Li Renpu
    OPTOELECTRONICS LETTERS, 2022, 18 (02) : 91 - 96
  • [2] Roll angle of autocollimator measurement method based on hollow cube corner reflector
    LIU Yu
    ZHEN Yan
    XIE Linyi
    WANG Wei
    YANG Jiawen
    LI Renpu
    Optoelectronics Letters, 2022, 18 (02) : 91 - 96
  • [3] Roll angle of autocollimator measurement method based on hollow cube corner reflector
    Yu Liu
    Yan Zhen
    Linyi Xie
    Wei Wang
    Jiawen Yang
    Renpu Li
    Optoelectronics Letters, 2022, 18 : 91 - 96
  • [4] A Method Based on Arc Template to Detect Right-angle Corner
    Li Jing
    Chen Huaimin
    Duan Xiaojun
    Duan Biao
    Ru Yi
    2018 INTERNATIONAL CONFERENCE ON ARTIFICIAL INTELLIGENCE AND BIG DATA (ICAIBD), 2018, : 179 - 184
  • [5] A new retroreflector named ?quatriplan? based on the corner-cube principle, for advanced ultrasonic telemetry applications
    Ploix, Marie-Aude
    Gueudre, Cecile
    Corneloup, Gilles
    Baque, Francois
    ULTRASONICS, 2023, 132
  • [6] Roll angle autocollimator measurement method based on a cylindrical cube-corner reflector with a high resolution and large range
    Li, Renpu
    Xie, Linyi
    Zhen, Yan
    Xiao, Han
    Wang, Wei
    Guo, Junqi
    Konyakhin, Igor
    Nikitin, Mikhail
    Yu, Xiaolong
    OPTICS EXPRESS, 2022, 30 (05) : 7147 - 7161
  • [7] Calibration of Angle of Incidence of Ellipsometer by Autocollimator-based Method
    Liu, Wende
    Chen, Chi
    Fan, Qiming
    Chu, Chu
    OPTICAL MEASUREMENT TECHNOLOGY AND INSTRUMENTATION, 2016, 10155
  • [8] TEST METHOD FOR SINGLE INTERIOR RIGHT ANGLE IN A CORNER CUBE PRISM
    WENG Xingtao HE Ping’an LI Song WANG JiyaoWENG Xingtao
    Geo-Spatial Information Science, 2000, (03) : 73 - 75
  • [9] A method to diverge reflected beam uniformly using cube-corner retroreflector array with dihedral angle tolerances
    Zhou Xiao-Feng
    Qi Zu-Min
    Luo Xiang-Qian
    Liu Chang-An
    Zhu Jian-Hui
    Wang Ze-Hua
    Zhang Yi
    Zi Yan-Yong
    ACTA PHYSICA SINICA, 2017, 66 (08)
  • [10] CORRECTING FOR X-RAY-ENERGY CALIBRATION ERROR CAUSED BY MISALIGNMENT OF A RIGHT-ANGLE LINKAGE MONOCHROMATOR
    KIM, KH
    BELL, MI
    DOZIER, CM
    FREITAG, RK
    BOULDIN, CE
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1991, 62 (04): : 982 - 985