Spatial Characteristics of Spectral Intensity of Laser Induced Plasma

被引:0
|
作者
Sun C. [1 ,2 ]
Bu Y. [1 ,2 ]
Wang Y. [1 ,2 ]
Wu F. [1 ,2 ]
Cai Y. [1 ,3 ]
Wang X. [1 ,2 ]
机构
[1] Laboratory of Information Optics and Opto-Electronic Technology, Shanghai Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Shanghai
[2] Center of Materials Science and Optoelectronics Engineering, University of Chinese Academy of Sciences, Beijing
[3] Department of Physics, Shanghai University, Shanghai
来源
Guangxue Xuebao/Acta Optica Sinica | 2021年 / 41卷 / 21期
关键词
Characteristic spectral line; Laser-induced breakdown spectroscopy; Optimal detection angle; Optimal incidence angle; Plasma; Spatial distribution; Spectroscopy;
D O I
10.3788/AOS202141.2130001
中图分类号
学科分类号
摘要
Laser wavelength and incident angle are important factors that influence the spatial distribution and spectral intensity of laser-induced plasma. Based on the fluid dynamics and SAHA equation, this paper simulates the two-dimensional spatial distribution of laser-induced plasma, calculates the spatial distribution of radiation spectrum of the excited plasma, and studies the influences of laser incident angle, wavelength and other parameters on the spatial distribution characteristics of plasma characteristic spectral lines. The research results show that that 0° is the best incident angle for 1064 nm laser with different delay conditions. When the incident angle is 0°, the excited plasma radiation has stronger spectral signals at different detection angles. Moreover, the optimal detection angles are ±41°, ±11° and ±12° at 100 ns, 500 ns and 1000 ns delay conditions, respectively. For different wavelengths, when the delay time is 100 ns or 500 ns and the laser is incident at 0°, the intensity of plasma spectrum excited by 1064 nm laser at each detection angle is stronger compared that by a short wavelength laser. When the time delay is 100 ns, the intensity of the plasma spectrum excited by the 1064 nm laser at the optimal detection angle is approximately twice that by 532 nm or 266 nm laser at each optimal detection angle. With the decrease of the absolute value of the detection angle, the spectral intensity of the plasma radiation first increases and then decreases at the optimal detection angle. The simulation results are verified by the laser induced breakdown spectroscopy experimental results with incident wavelengths of 532 nm and 1064 nm, respectively. © 2021, Chinese Lasers Press. All right reserved.
引用
收藏
相关论文
共 21 条
  • [1] Sha W, Li J T, Lu C P., Quantitative analysis of Mn in soil based on laser-induced breakdown spectroscopy optimization, Chinese Journal of Lasers, 47, 5, (2020)
  • [2] Babos D V, Cruz-Conesa A, Pereira-Filho E R, Et al., Direct determination of Al and Pb in waste printed circuit boards (PCB) by laser-induced breakdown spectroscopy (LIBS): evaluation of calibration strategies and economic-environmental questions, Journal of Hazardous Materials, 399, (2020)
  • [3] Moon Y, Han J H, Choi J H, Et al., Mapping of cutaneous melanoma by femtosecond laser-induced breakdown spectroscopy, Journal of Biomedical Optics, 24, 3, pp. 1-6, (2018)
  • [4] Li J Y, Chen K, Chen G F, Et al., Quantitative analysis of magnesium and titanium elements in aluminum alloy by handheld laser-induced breakdown spectroscopy, Laser & Optoelectronics Progress, 56, 2, (2019)
  • [5] Lawley C J M, Somers A M, Kjarsgaard B A., Rapid geochemical imaging of rocks and minerals with handheld laser induced breakdown spectroscopy (LIBS), Journal of Geochemical Exploration, 222, (2021)
  • [6] Xu X J, Wang X S, Li A Z, Et al., Fast classification of tea varieties based on laser-induced breakdown spectroscopy, Chinese Journal of Lasers, 46, 3, (2019)
  • [7] Nosrati Y, Tavassoli S H, Hassanimatin M M, Et al., Study of material ablation and plasma radiation in double-pulse laser induced breakdown spectroscopy at different delay times: modeling and numerical simulation, Physics of Plasmas, 27, 2, (2020)
  • [8] Shabanov S V, Gornushkin I B., Two-dimensional axisymmetric models of laser induced plasmas relevant to laser induced breakdown spectroscopy, Spectrochimica Acta Part B: Atomic Spectroscopy, 100, pp. 147-172, (2014)
  • [9] Aghaei M, Mehrabian S, Tavassoli S H., Simulation of nanosecond pulsed laser ablation of copper samples: a focus on laser induced plasma radiation, Journal of Applied Physics, 104, 5, (2008)
  • [10] Lee Y T, More R M., An electron conductivity model for dense plasmas, Physics of Fluids, 27, 5, pp. 1273-1286, (1984)