ANALYZER SYSTEM FOR FIELD-EMISSION ENERGY-DISTRIBUTION (FEED) MEASUREMENTS

被引:5
|
作者
KEMPIN, HF [1 ]
KLAPPER, K [1 ]
ERTL, G [1 ]
机构
[1] UNIV MUNICH,INST PHYS CHEM,D-8000 MUNICH 2,FED REP GER
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 1978年 / 49卷 / 09期
关键词
D O I
10.1063/1.1135569
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A newly designed analyzer system for measuring the energy distribution of field emitted electrons (FEED) is presented. A Kuyatt-Simpson type lens system is combined with a two-stage 127 degree analyzer. A resolution of 15 meV is obtained with a transmission energy of 0. 5 eV.