A SiPM-Readout ASIC for SPECT Applications

被引:25
|
作者
Trigilio, Paolo [1 ,2 ,3 ]
Busca, Paolo [1 ,2 ,4 ]
Quaglia, Riccardo [1 ,2 ,5 ]
Occhipinti, Michele [1 ,2 ,3 ]
Fiorini, Carlo [1 ,2 ]
机构
[1] Politecn Milan, Dipartimento Elettron Informaz & Bioingn, I-20133 Milan, Italy
[2] INFN, Sez Elettron, I-20133 Milan, Italy
[3] XGlab Srl, I-20134 Milan, Italy
[4] ESRF European Synchrotron, Instrumentat Serv & Dev Div, F-38000 Grenoble, France
[5] Swiss Ctr Elect & Microtechnol, Embedded Vis Syst, CH-2002 Neuchatel, Switzerland
关键词
Application specific integrated circuit (ASIC); silicon photomultiplier (SiPM); SiPM front-end; SiPM readout; single photon emission computed tomography (SPECT); SPECT/MRI;
D O I
10.1109/TRPMS.2018.2856201
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this paper, we present ANGUS, an application specific integrated circuit (ASIC) for silicon photomultiplier (SiPM) readout designed for low-energy gamma imaging (140 keV), specifically targeting clinical single photon emission computed tomography (SPECT) applications based on Angertype detectors. The chip is composed of 36 channels, each featuring a current conveyor input followed by an RC filter with a time constant programmable from 200 ns up to 10 mu s, a baseline holder and a peak stretcher stage. Multichannel readout is performed with a polling strategy thanks to two analog multiplexers. The chip has been designed to cope with input capacitance values of several nanofarad, like the ones featured by the large-sized SiPM pixels (8 mm x 8 mm) of the application. Despite being conceived for operation with slow scintillator crystals like cesium-iodide (CsI), this ASIC can successfully manage the readout of faster crystals like lutetium-yttrium oxyorthosilicate. Measurements of the ASIC carried out on a fully equipped SPECT detection module have shown an energy resolution of 13.7% at the 140-keV line of 99mTc, for a 5 cm x 5 cm gamma camera consisting of a continuous CsI crystal and a SiPM detection plane cooled at 0 degrees C.
引用
收藏
页码:404 / 410
页数:7
相关论文
共 50 条
  • [1] ASIC development for SiPM readout
    Corsi, Francesco
    Foresta, Maurizio
    Marzocca, Cristoforo
    Matarrese, Gianvito
    Del Guerra, Alberto
    JOURNAL OF INSTRUMENTATION, 2009, 4
  • [2] An ASIC for SiPM/MPPC Readout
    Meier, Dirk
    Mikkelsen, Sindre
    Talebi, Jahanzad
    Azman, Suleyman
    Maehlum, Gunnar
    Patt, Bradley E.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 1653 - 1657
  • [3] A readout ASIC for SPECT
    Pettersen, DM
    Mikkelsen, S
    Talebi, J
    Meier, D
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2204 - 2210
  • [4] A readout ASIC for SPECT
    Pettersen, DA
    Mikkelsen, S
    Talebi, J
    Meier, D
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (03) : 764 - 771
  • [5] Low-power SiPM readout BETA ASIC for space applications
    Sanmukh, Anand
    Gomez, Sergio
    Comerma, Albert
    Mauricio, Joan
    Manera, Rafel
    Sanuy, Andreu
    Guberman, Daniel
    Catala, Roger
    Espinya, Albert
    Orta, Marina
    de la Torre, Oscar
    Gascon, David
    NUCLEAR SCIENCE AND TECHNIQUES, 2024, 35 (03)
  • [6] Low-power SiPM readout BETA ASIC for space applications
    Anand Sanmukh
    Sergio Gómez
    Albert Comerma
    Joan Mauricio
    Rafel Manera
    Andreu Sanuy
    Daniel Guberman
    Roger Catala
    Albert Espinya
    Marina Orta
    Oscar de la Torre
    David Gascon
    NuclearScienceandTechniques, 2024, 35 (03) : 155 - 171
  • [7] Front End ASIC design for SiPM readout
    Comerma, A.
    Gascon, D.
    Garrido, L.
    Delgado, C.
    Marin, J.
    Perez, J. M.
    Martinez, G.
    Freixas, L.
    JOURNAL OF INSTRUMENTATION, 2013, 8
  • [8] PACIFIC: SiPM readout ASIC for LHCb upgrade
    Mazorra de Cos, Jose
    Chanal, Herve
    Montells, Albert Comerma
    Gascon Fora, David
    Gomez Fernandez, Sergio
    Han, Xiaoxue
    Pillet, Nicolas
    Vandaelle, Richard
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 912 : 354 - 358
  • [9] Characterization results and first applications of KLauS - an ASIC for SiPM charge and fast discrimination readout
    Briggl, Konrad
    Shen, Wei
    Harion, Tobias
    Hagdorn, Rene
    Eckert, Patrick
    Klapdor-Kleingrothaus, Thorwald
    Dorn, Markus
    Schultz-Coulon, Hans-Christian
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [10] Performance studies of a SiPM-readout system with a pico-second timing chip
    Xia, Xin
    Du, Dejing
    Jiang, Xiaoshan
    Liu, Yong
    Lu, Bo
    Lyu, Junguang
    Qi, Baohua
    Ruan, Manqi
    Yan, Xiongbo
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2024, 1064