SURFACE-ANALYSIS OF SURFACE-PASSIVATED INTRAOCULAR LENSES

被引:14
|
作者
KOCH, DD
SAMUELSON, SW
DIMONIE, V
机构
[1] Cullen Eye Institute, Houston, Texas
[2] Emulsion Polymers Institute, Lehigh University, Bethlehem, Pennsylvania
来源
关键词
CONTACT ANGLE; POLYFLUOROCARBON COATING; POLY(METHYL METHACRYLATE); POLYVINYLPYRROLIDONE GRAFTING; SMART (SELECTIVE MOLECULAR ALTERATION AND REALIGNMENT TECHNOLOGY); SURFACE ENERGY; SURFACE MODIFICATION; SURFACE PASSIVATION; SURFACE TENSION;
D O I
10.1016/S0886-3350(13)80242-8
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
We analyzed the surfaces of six different types of poly(methyl methacrylate) intraocular lenses using electron spectroscopy for chemical analysis (ESCA), static secondary ion mass spectroscopy (SIMS), and two techniques for determining surface energies. The tested lenses were (1) Ioptex lathe-cut surface-passivated, (2) Ioptex lathe-cut, (3) another manufacturer's lathe-cut, (4) cast-molded, (5) polyfluorocarbon-coated, and (6) polyvinylpyrrolidone-grafted. The ESCA testing revealed marked differences in the chemical composition of the surfaces of the polyfluorocarbon-coated and polyvinypyrrolidone-grafted lenses compared to the other four. Minute amounts of silicon were present on the cast-molded and surface-passivated lenses. Static SIMS testing revealed the presence of siloxane contaminants on the three lathe-cut and the cast-molded lenses; a statistically smaller amount of siloxane was present on the surface-passivated lenses. Contact angle and surface energy analysis revealed statistically lower surface energies for the polyfluorocarbon-coated materials and statistically higher surface energies for the polyvinylpyrrolidone-grafted; we found no differences in the contact angles and surface energies among the surface-passivated, lathe-cut, and cast-molded lenses.
引用
收藏
页码:131 / 138
页数:8
相关论文
共 50 条
  • [1] Analysis of electron transport in surface-passivated nanocrystalline porous silicon
    Koijma, A
    Koshida, N
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2003, 42 (4B): : 2395 - 2398
  • [2] Electrical Properties of Surface-Passivated GaAs Nanowires
    Lee, Rochelle S.
    Kim, Tae Kyum
    Lee, Sang Won
    Cho, Kyu Yeon
    Choi, Jong Hyun
    Kim, Mi Yeong
    Shin, Jae Cheol
    APPLIED SCIENCE AND CONVERGENCE TECHNOLOGY, 2018, 27 (06): : 166 - 168
  • [3] ANALYSIS OF SURFACE CONTAMINANTS ON INTRAOCULAR LENSES
    RATNER, BD
    ARCHIVES OF OPHTHALMOLOGY, 1983, 101 (09) : 1434 - 1438
  • [4] Surface-passivated high-resistivity silicon substrates for RFICs
    Rong, B
    Burghartz, JN
    Nanver, LK
    Rejaei, B
    van der Zwan, M
    IEEE ELECTRON DEVICE LETTERS, 2004, 25 (04) : 176 - 178
  • [5] Spherical silicon photovoltaic devices with surface-passivated shallow emitters
    Gharghi, M.
    Sivoththaman, S.
    SEMICONDUCTOR SCIENCE AND TECHNOLOGY, 2008, 23 (10)
  • [6] Synthesis of size-selected, surface-passivated InP nanocrystals
    Guzelian, AA
    Katari, JEB
    Kadavanich, AV
    Banin, U
    Hamad, K
    Juban, E
    Alivisatos, AP
    Wolters, RH
    Arnold, CC
    Heath, JR
    JOURNAL OF PHYSICAL CHEMISTRY, 1996, 100 (17): : 7212 - 7219
  • [7] Epitaxial growth of surface-passivated core-shell ferrierite
    Han, Sungmin
    Shilpa, Kumari
    Mallette, Adam J.
    Li, Yuejin
    Hoke, Jeffrey B.
    Rimer, Jeffrey D.
    JOURNAL OF CRYSTAL GROWTH, 2023, 603
  • [8] Polarized photoluminescence from surface-passivated lead sulfide nanocrystals
    Fernée, MJ
    Warner, J
    Watt, A
    Cooper, S
    Heckenberg, NR
    Rubinsztein-Dunlop, H
    NANOTECHNOLOGY, 2004, 15 (01) : 16 - 22
  • [9] Surface Modification of Intraocular Lenses
    Huang Qi
    Cheng George Pak-Man
    Chiu Kin
    Wang Gui-Qin
    中华医学杂志英文版, 2016, 129 (02) : 206 - 214
  • [10] SURFACE CHARACTERIZATION OF INTRAOCULAR LENSES
    STICKLE, WF
    REICHLMAIER, S
    SOBOL, PE
    MCKEOWN, P
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1993, 206 : 119 - POLY