FLOW-INJECTION DETERMINATION OF 9,10-PHENANTHRENEQUINONE WITH CATALYTIC PHOTOMETRIC DETECTION

被引:5
|
作者
KIBA, N
SUZUKI, H
GOTO, E
FURUSAWA, M
机构
[1] Department of Applied Chemistry and Biotechnology, Faculty of Engineering, Yamanashi University, Kofu
关键词
D O I
10.1016/0039-9140(93)80252-M
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A catalytic photometric method with a flow-injection system is described for the determination of 9,10-phenanthrenequinone. It is based on the catalytic effect of 9,10-phenanthrenequinone on the redox reaction of 1,2-dinitrobenzene with formaldehyde under alkaline conditions. 9,10-Phenanthrenequinone at the 5.0 x 10(-8)-5.0 x 10(-6)M level can be determined at a rate of 20 samples/hr. The detection limit is 1.0 x 10(-8)M (40 pg in a 10-mul injection).
引用
收藏
页码:405 / 407