A simple and fast HPLC method for the determination of glycols is described. It is characterized by a reversed-phase separation using water as eluent and a refractive index detection. The method was applied to investigate the biodegradation of glycols in a laboratory activated sludge plant and to determine the content of glycols or alcohols in detergents. The detection limit is 4 mg/l ethylene glycol or propylene glycol in an aqueous sample.