DOUBLED MULTIPLEXING IN FOURIER-TRANSFORM RAMAN-SPECTROSCOPY

被引:6
|
作者
RADZISZEWSKI, JG
MICHL, J
机构
[1] Univ of Texas at Austin, , TX
关键词
D O I
10.1366/0003702904086254
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The use of a Chevron filter for the rejection of light at the 1064-nm excitation wavelength permits the simultaneous detection of Stokes and anti-Stokes Raman lines (doubled multiplexing). This enhances spectrometric and photometric accuracy and permits a measurement of the true vibrational temperature of the sample. Down to Raman shifts as low as 40 cm-1, near-IR Raman spectra recorded on an optimized Nicolet SXR 60 instrument are distinctly superior to previously reported spectra of similar samples recorded on FT or, more slowly, on dispersive instruments.
引用
收藏
页码:414 / 418
页数:5