共 50 条
SCREENING FOR DRUGS OF ABUSE BY SYVA EMIT(R) II REAGENTS ON THE BM/HITACHI 717 ANALYZER
被引:0
|作者:
TSO, G
WU, AHB
KUNTZ, D
WONG, SS
机构:
[1] UNIV TEXAS,SCH MED,DEPT PATHOL & LAB MED,HOUSTON,TX 77030
[2] HERMANN HOSP,DEPT LAB SERV,HOUSTON,TX 77030
[3] NW TOXICOL,SALT LAKE CITY,UT 84124
关键词:
D O I:
暂无
中图分类号:
R194 [卫生标准、卫生检查、医药管理];
学科分类号:
摘要:
The performance of Syva Emit(R) II reagents on the BM/Hitachi 717 analyzer for the screening of phencyclidine (PCP) and cannabinoids (THC) was evaluated. All procedures stipulated by the manufacturer for the semi-quantitative analysis were followed. Within-run and between-run precision studies showed that the coefficient of variation for both assays was below 0.97% for all concentrations tested. in addition there was no detectable sample carryover. These new products provided accurate drug-of-abuse screening at cutoff limits of 25 ng/mL for PCP and 20, 50 and 100 ng/mL for THC, depending on the Emit II reagents used. When compared to the Emit(R) d.a.u.(TM) assays the on Syva ETS(R) analyzer or the BM/Hitachi 717, the Emit II reagents had improved accuracy and precision. The use of the Emit II reagents on BM/Hitachi 717 instrument allows for screening of large batches of samples for drugs-of-abuse.
引用
收藏
页码:256 / 268
页数:13
相关论文