AVOIDING ATRIAL UNDERSENSING BY ASSESSMENT OF P-WAVE AMPLITUDE HISTOGRAM DATA

被引:19
|
作者
BOUTE, W
ALBERS, BA
GIELE, V
机构
[1] Vitatron Medical B.V, Velp
来源
关键词
ATRIAL UNDERSENSING; DUAL CHAMBER PACEMAKER; HISTOGRAM;
D O I
10.1111/j.1540-8159.1994.tb03767.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Reliable sensing of the P wave is an essential requirement for the appropriate functioning of any device that uses atrial tracking to provide AV synchrony. However, a single measurement of the P wave amplitude, either at implantation or during follow-up, may not be a reliable reflection of the P wave amplitudes that occur during daily life. The P wave amplitude histogram is a new feature that automatically measures the P wave amplitude at regular intervals and provides the distribution of these measurements. This enables the assessment of the smallest P wave amplitudes that occur. Two populations were studied: 104 patients with a fixed atrial lead and a DDDR pacemaker and 100 patients with a single pass VDD lead and a VDD pacemaker. Both pacemakers incorporate the P wave amplitude histogram feature. Data in the P wave amplitude histogram were compared with a single measurement of the P wave amplitude at each follow-up. Programming of a 100% safety margin based on a single measurement of the P wave amplitude provided reliable atrial sensing in only 72% and 43% of the patients of both populations, respectively. Data continued in the P wave amplitude histogram may be a useful adjunct for the optimal programming of atrial sensitivity.
引用
收藏
页码:1878 / 1882
页数:5
相关论文
共 50 条
  • [1] INFLUENCE OF ATRIAL MASS ON AMPLITUDE AND DURATION OF P-WAVE
    COKKINOS, DV
    ZAMALLOA, O
    CABRERA, R
    FERRER, MI
    LEACHMAN, RD
    CHEST, 1972, 61 (04) : 336 - &
  • [2] APPARENT P-WAVE UNDERSENSING IN A DDD PACEMAKER POST EXERCISE
    VANGELDER, BM
    VANMECHELEN, R
    DENDULK, K
    ELGAMAL, MIH
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11): : 1651 - 1656
  • [3] UNDERSENSING OF P-WAVES IN THE PRESENCE OF AN ADEQUATE P-WAVE DUE TO AUTOMATIC POSTVENTRICULAR ATRIAL REFRACTORY PERIOD EXTENSION
    WILSON, JH
    LATTNER, S
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (11): : 1729 - 1732
  • [4] ORTHOGONAL ATRIAL ELECTROGRAMS - HIGH AMPLITUDE, DISCRIMINATE P-WAVE SENSING
    GOLDREYER, BN
    BRUESKE, R
    KNUDSON, M
    CANNOM, DS
    WYMAN, MG
    CIRCULATION, 1982, 66 (04) : 375 - 375
  • [5] Risk assessment for atrial fibrillation: Enter the P-wave
    Magnani, Jared W.
    HEART RHYTHM, 2015, 12 (09) : 1896 - 1897
  • [6] What is the cause of P-wave undersensing in this CRT-D device?
    Schroemges, J.
    Bracke, F. A. L. E.
    van Gelder, B. M.
    NETHERLANDS HEART JOURNAL, 2018, 26 (7-8) : 409 - +
  • [7] What is the cause of P-wave undersensing in this CRT-D device?
    J. Schroemges
    F. A. L. E. Bracke
    B. M. van Gelder
    Netherlands Heart Journal, 2018, 26 : 409 - 410
  • [8] Assessment of P-wave dispersion in children with atrial septal aneurysm
    Arslan, Derya
    Cimen, Derya
    Guvenc, Osman
    Oran, Bulent
    Yilmaz, Fatma Hilal
    CARDIOLOGY IN THE YOUNG, 2014, 24 (05) : 918 - 922
  • [9] P-WAVE AMPLITUDE VARIABILITY AT NORSAR
    BERTEUSSEN, KA
    JOURNAL OF GEOPHYSICS-ZEITSCHRIFT FUR GEOPHYSIK, 1975, 41 (06): : 595 - 613
  • [10] Left Atrial Hypertrophy Increases P-Wave Terminal Force Through Amplitude but not Duration
    Loewe, Axel
    Andlauer, Robin
    Platonov, Pyotr G.
    Doessel, Olaf
    Seemann, Gunnar
    2016 COMPUTING IN CARDIOLOGY CONFERENCE (CINC), VOL 43, 2016, 43 : 1 - 4