THE EFFECT OF HIGH-PERFORMANCE GRADIENTS ON FAST GRADIENT-ECHO IMAGING

被引:40
|
作者
REEDER, SB [1 ]
MCVEIGH, ER [1 ]
机构
[1] JOHNS HOPKINS UNIV,SCH MED,DEPT BIOMED ENGN & RADIOL,MED IMAGING LAB,BALTIMORE,MD 21205
关键词
OPTIMIZATION; MRI; SNR; GRADIENT COILS;
D O I
10.1002/mrm.1910320510
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The effect of gradient system performance on segmented k-space gradient echo imaging is presented. Three cases were investigated. First, an ideal system that has infinite slew rates and unlimited maximum gradient strengths was considered. Second, a ''high speed'' imaging system (2.3 (G/cm), 23 (G/cm)/ms) was considered. These two cases were compared with a ''conventional'' imaging system (1 (G/cm), 1.67 (G/cm)/ms). It was found that substantial increases in SNR can be achieved (approximate to 45%) by using high speed versus a conventional gradient system, for a TR of 6 ms. For trapezoidal gradient waveforms, there exists an optimum maximum gradient strength for a given slew rate, and any increase in gradient strength above this optimum will not be utilized by an optimized sequence. These studies have shown that increasing TR without decreasing the bandwidth is not a good way to increase SNR for constant scan time.
引用
收藏
页码:612 / 621
页数:10
相关论文
共 50 条