INTERFERENCE-FREE GLUCOSE SENSOR BASED ON GLUCOSE-OXIDASE IMMOBILIZED IN AN OVEROXIDIZED NONCONDUCTING POLYPYRROLE FILM

被引:105
|
作者
CENTONZE, D [1 ]
GUERRIERI, A [1 ]
MALITESTA, C [1 ]
PALMISANO, F [1 ]
ZAMBONIN, PG [1 ]
机构
[1] UNIV BARI,DIPARTIMENTO CHIM,CHIM ANALIT LAB,4 TRAV 200 RE DAVID,I-70126 BARI,ITALY
来源
关键词
D O I
10.1007/BF00321865
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An amperometric glucose sensor is described; it is based on glucose oxidase immobilized in an overoxidized non-conducting polypyrrole membrane. The overoxidation process of polypyrrole produces a permselective, antifouling membrane capable of rejecting ascorbate, urate, acetaminophen and cysteine, as well as proteins and other surface active components typically present in serum. The amperometric assay of glucose in serum correlates well with an established routine procedure based on an enzymatic-colorimetric method.
引用
收藏
页码:729 / 733
页数:5
相关论文
共 50 条