Comparing Radiation Dose from Conventional Fluoroscopy to Intraoperative Cone Beam CT (O-arm) during Percutaneous Lesioning Procedures of the Gasserian Ganglion

被引:8
|
作者
Desai, Sohum [1 ]
Patel, Vishal J. [1 ]
Lall, Rishi R. [1 ]
Branch, Daniel [1 ]
Patel, Achal P. [1 ]
Allison, Randall Z. [1 ]
Paulson, David [2 ]
Ortega-Barnett, Juan R. [1 ]
机构
[1] Univ Texas Med Branch Galveston, Div Neurosurg, Galveston, TX 77555 USA
[2] Swedish Neurosci Inst, Neurosurg, Seattle, WA USA
来源
CUREUS | 2015年 / 7卷 / 10期
关键词
trigeminal neuralgia; cone beam ct; radiation dose; fluoroscopy; percutaneous trigeminal rhizotomy;
D O I
10.7759/cureus.345
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: The use of intraoperative CT-guidance during the percutaneous treatment of trigeminal neuralgia has become increasingly popular due to the greater ease of foramen ovale cannulation and decreased procedure times. Concerns regarding radiation dose to the patient, however, remain unaddressed. We sought to compare the emitted radiation dose from fluoroscopy with intraoperative CT for these procedures. Methods: A retrospective review of percutaneous lesioning procedures for trigeminal neuralgia performed between 2010 until 2012 at our institution was conducted and radiation doses to the patient were recorded. We subsequently simulated four separate percutaneous trigeminal rhizotomies using the O-arm intraoperative CT (Medtronics, Minneapolis, MN, USA) to cannulate the foramen ovale bilaterally in two formalin-fixed cadaver heads. Results: Seventeen successful percutaneous treatments for trigeminal neuralgia were performed during the study period. Eleven procedures containing complete records were included in the final analysis. For procedures using fluoroscopy, the mean dosage was 15.2 mGys (range: 1.15 - 47.95, 95% CI 7.34 - 22.99). Radiation dosage from the O-arm imaging system was 16.55 mGy for all four cases. An unequal variance t-test did not reach statistical significance (p=0.42). Conclusions: We did not observe a significant difference in radiation dose delivered to subjects when comparing CT-guided foramen ovale cannulation relative to fluoroscopy for percutaneous lesioning of the Gasserian ganglion. Additional study is required under operational settings.
引用
收藏
页数:6
相关论文
共 16 条
  • [1] Intraoperative Cone Beam-Computed Tomography With Navigation (O-ARM) Versus Conventional Fluoroscopy (C-ARM)
    Tabaraee, Ehsan
    Gibson, Anthony G.
    Karahalios, Dean G.
    Potts, Eric A.
    Mobasser, Jean-Pierre
    Burch, Shane
    SPINE, 2013, 38 (22) : 1953 - 1958
  • [2] Intraoperative Cone-Beam CT Image Quality and Dose with a Slot Collimator On the O-Arm
    Zhang, X.
    Wu, P.
    Huang, Y.
    Uneri, A.
    Jones, C.
    Helm, P.
    Lo, S.
    Zbijewski, W.
    Siewerdsen, J.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [3] Patient radiation dose during CT fluoroscopy guided percutaneous interventional procedures
    Daly, BD
    Regits, W
    Pham, CN
    White, DE
    Hartman, EM
    Westerman, BR
    RADIOLOGY, 1999, 213P : 285 - 285
  • [4] Implementation of Intraoperative Cone-Beam Computed Tomography (O-arm) for Stereotactic Imaging During Deep Brain Stimulation Procedures
    Holewijn, Rozemarije A.
    Bot, Maarten
    van den Munckhof, Pepijn
    Schuurman, P. Richard
    OPERATIVE NEUROSURGERY, 2020, 19 (03) : E224 - E229
  • [5] A method to characterize the radiation output from a cone beam O-arm using a device for dose and dose profile scanning measurement
    Herrnsdorf, Lars
    Soderberg, Marcus
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [6] Intraoperative cone-beam and slot-beam CT: 3D image quality and dose with a slot collimator on the O-arm imaging system
    Zhang, Xiaoxuan
    Zbijewski, Wojciech
    Huang, Yixuan
    Uneri, Ali
    Jones, Craig K.
    Lo, Sheng-Fu L.
    Witham, Timothy F.
    Luciano, Mark
    Anderson, William Stanley
    Helm, Patrick A.
    Siewerdsen, Jeffrey H.
    MEDICAL PHYSICS, 2021, 48 (11) : 6800 - 6809
  • [7] Surgivisio® and O-arm®O2 cone beam CT mobile systems for guidance of lumbar spine surgery: Comparison of patient radiation dose
    Rousseau, Julia
    Dreuil, Serge
    Bassinet, Celine
    Cao, Sophie
    Elleaume, Helene
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 85 : 192 - 199
  • [8] Optimization of Radiation Exposure and Image Quality of the Cone-beam O-arm Intraoperative Imaging System in Spinal Surgery
    Abul-Kasim, Kasim
    Soderberg, Marcus
    Selariu, Eufrozina
    Gunnarsson, Mikael
    Kherad, Mehrsa
    Ohlin, Acke
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2012, 25 (01): : 52 - 58
  • [9] Estimation of organ and effective dose to the patient during spinal surgery with a cone-beam O-arm system
    Soderberg, Marcus
    Abul-Kasim, Kasim
    Ohlin, Acke
    Gunnarsson, Mikael
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [10] Low-Dose Radiation 3D Intraoperative Imaging How Low Can We Go? An O-Arm, CT Scan, Cadaveric Study
    Sarwahi, Vishal
    Payares, Monica
    Wendolowski, Stephen
    Maguire, Kathleen
    Thornhill, Beverly
    Lo, Yungtai
    Amaral, Terry D.
    SPINE, 2017, 42 (22) : E1311 - E1317