摘要
This paper presents a dosimetric study concerning the system named "Papillon 50" used in the department of radiotherapy of the Centre Antoine-Lacassagne, Nice, France. The machine provides a 50 kVp X-ray beam, currently used to treat rectal cancers. The system can be mounted with various applicators of different diameters or shapes. These applicators can be fixed over the main rod tube of the unit in order to deliver the prescribed absorbed dose into the tumor with an optimal distribution. We have analyzed depth dose curves and dose profiles for the naked tube and for a set of three applicators. Dose measurements were made with an ionization chamber (PTV type 23342) and Gafchromic films (EBT2). We have also compared the measurements with simulations performed using the Monte Carlo code PENELOPE. Simulations were performed with a detailed geometrical description of the experimental setup and with enough statistics. Results of simulations are made in accordance with experimental measurements and provide an accurate evaluation of the dose delivered. The depths of the 50% isodose in water for the various applicators are 4.0, 6.0, 6.6 and 7.1 mm. The Monte Carlo PENELOPE simulations are in accordance with the measurements for a 50 kV X-ray system. Simulations are able to confirm the measurements provided by Gafchromic films or ionization chambers. Results also demonstrate that Monte Carlo simulations could be helpful to validate the future applicators designed for other localizations such as breast or skin cancers. Furthermore, Monte Carlo simulations could be a reliable alternative for a rapid evaluation of the dose delivered by such a system that uses multiple designs of applicators. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:609 / 617
页数:9
相关论文
共 50 条
  • [1] Monte Carlo simulation of X-ray emission using the general-purpose code PENELOPE
    Llovet, X
    Fernández-Varea, JM
    Sempau, J
    Salvat, F
    SURFACE AND INTERFACE ANALYSIS, 2005, 37 (11) : 1054 - 1058
  • [2] Targeted Radiotherapy Using Contact X-ray Brachytherapy 50 kV
    Gerard, Jean-Pierre
    Myint, Arthur Sun
    Barbet, Nicolas
    Dejean, Catherine
    Thamphya, Brice
    Gal, Jocelyn
    Montagne, Lucile
    Vuong, Te
    CANCERS, 2022, 14 (05)
  • [3] Monte Carlo modelling of radiotherapy kV x-ray units
    Verhaegen, F
    Nahum, AE
    Van de Putte, S
    Namito, Y
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (07): : 1767 - 1789
  • [4] X-ray simulation with the Monte Carlo code PENELOPE. Application to Quality Control
    Pozuelo, F.
    Gallardo, S.
    Querol, A.
    Verdu, G.
    Rodenas, J.
    2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 5777 - 5780
  • [5] Evaluation of Dose Distribution by Categories in X-Ray CT Room Using Monte Carlo Simulation
    Fujibuchi, T.
    Ueno, D.
    MEDICAL PHYSICS, 2019, 46 (06) : E239 - E239
  • [6] Monte Carlo (PENELOPE code) study of the x-ray beams from SL Linacs (Elekta)
    Bramoullé, C
    Husson, F
    Manens, JP
    PHYSICA MEDICA, 2000, 16 (03) : 107 - 115
  • [7] DETERMINATION OF DOSE DISTRIBUTION ON LUNG SIMULATOR OBJECT WITH MONTE CARLO SIMULATION USING CODE PENELOPE
    Apaza, Danny G.
    Chahuara, Ruben D.
    Sellerico, Pedro P.
    Vega, Jose L.
    MOMENTO-REVISTA DE FISICA, 2019, (59): : 1 - 10
  • [8] Monte Carlo study of dose deposition in kilovoltage X-ray radiotherapy using gold as dose enhancer
    Spiga, J.
    Pellicioli, P.
    Duffy, J.
    Manger, S.
    Bravin, A.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S755 - S755
  • [9] Monte Carlo Simulation of X-Ray Spectra in Mammography and Contrast-Enhanced Digital Mammography Using the Code PENELOPE
    Cunha, Diego M.
    Tomal, Alessandra
    Poletti, Martin E.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (02) : 495 - 502
  • [10] Combined x-ray/electron/optical Monte Carlo code based on PENELOPE and DETECT-II
    Badano, A
    Sempau, J
    Boswell, JS
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 870 - 876