DISMOUNTABLE SAMPLE HOLDER APPARATUS FOR RAPID THERMAL CONDUCTIVITY MEASUREMENTS BASED ON CRYOCOOLER

被引:7
|
作者
Liu, Huiming [1 ]
Xu, Dong [1 ]
Xu, Peng [1 ]
Huang, Rongjin [1 ]
Xu, Xiangdong [1 ]
Li, Laifeng [1 ]
Gong, Linghui [1 ]
机构
[1] Chinese Acad Sci, Key Lab Cryogen, Tech Inst Phys & Chem, Beijing 100190, Peoples R China
关键词
thermal conductivity; cryocooler; dismountable sample test bar; thermal bridge; steady-state absolute method;
D O I
10.1063/1.4707062
中图分类号
O59 [应用物理学];
学科分类号
摘要
A novel apparatus, based on cryocooler, for rapid thermal conductivity measurements between 8 K and 300 K is presented. It consists of a removable sample test bar on which bulk samples can easily be mounted and then placed in the described measurement device. This fast mounting measurement system uses a standard steady-state absolute thermal conductivity measurement and allows for excellent thermal stability and mechanical vibration isolation from the cryocooler. The distinction of this system is rapid mounting and measurement of thermal conductivity with high accuracy and precision in data acquisition. In addition, this system allows for versatility in its use, such as the specific heat and the linear thermal expansion measurement. The design of this apparatus, measurement specification, and thermal conductivity of standard materials measured in this system are presented.
引用
收藏
页码:1363 / 1370
页数:8
相关论文
共 50 条
  • [21] Numerical analysis of sample dimensions in hot wire thermal conductivity measurements
    Cintra, JD
    dos Santos, WN
    JOURNAL OF THE EUROPEAN CERAMIC SOCIETY, 2000, 20 (11) : 1871 - 1875
  • [22] Cryogen-free apparatus for rapid thermal emissivity measurements at low temperature
    Shen, F. Z.
    Xu, D.
    Li, X.
    Liu, H. M.
    Li, L. F.
    27TH INTERNATIONAL CRYOGENICS ENGINEERING CONFERENCE AND INTERNATIONAL CRYOGENIC MATERIALS CONFERENCE 2018 (ICEC-ICMC 2018), 2019, 502
  • [23] Design and test of an apparatus for transient measurements of the thermal conductivity and volumetric heat capacity of solids
    Gather, Florian
    Gunes, Ekrem
    Klar, Peter J.
    TURKISH JOURNAL OF PHYSICS, 2019, 43 (01): : 37 - 50
  • [24] THERMAL CONDUCTIVITY OF METALS AT HIGH TEMPERATURES .1. DESCRIPTION OF THE APPARATUS AND MEASUREMENTS ON IRON
    ARMSTRONG, LD
    DAUPHINEE, TM
    CANADIAN JOURNAL OF RESEARCH SECTION A-PHYSICAL SCIENCES, 1947, 25 (06): : 357 - 374
  • [25] Test of an apparatus for thermal conductivity measurements of superconducting coil blocks and materials at cryogenic temperatures
    Broggi, F
    Rossi, L
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1996, 67 (09): : 3193 - 3200
  • [26] Test of an apparatus for thermal conductivity measurements of superconducting coil blocks and materials at cryogenic temperatures
    Universita' di Milano, Segrate, Italy
    Rev Sci Instrum, 9 (3193-3200):
  • [27] DEVELOPMENT OF A SINGLE-SIDED GUARDED HOT PLATE APPARATUS FOR THERMAL CONDUCTIVITY MEASUREMENTS
    Terzic, Marijana M.
    Milosevic, Nenad D.
    Stepanic, Nenad M.
    Petricevic, Slobodan J.
    THERMAL SCIENCE, 2016, 20 : S321 - S329
  • [28] Sample holder design for effective thermal conductivity measurement of pebble-bed using laser flash method
    Lee, Youngmin
    Ku, Duck Young
    Park, Yi-Hyun
    Ahn, Mu-Young
    Cho, Seungyon
    FUSION ENGINEERING AND DESIGN, 2017, 124 : 995 - 998
  • [29] Construction and preliminary testing of a guarded hot plate apparatus for thermal conductivity measurements at high temperatures
    Yang, Inseok
    Kim, Daeho
    Lee, Sanghyun
    INTERNATIONAL JOURNAL OF HEAT AND MASS TRANSFER, 2018, 122 : 1343 - 1352
  • [30] MEASUREMENTS OF THERMAL-CONDUCTIVITY OF MATERIALS USING A TRANSIENT TECHNIQUE .2. DESCRIPTION OF THE APPARATUS
    PIORKOWSKA, E
    GALESKI, A
    JOURNAL OF APPLIED PHYSICS, 1986, 60 (02) : 493 - 498